FSV: Remote reading of the information in the status bar?  FAQ overview
 
Question: Please advise me of a way/method (if applicable) of (remotely) reading the information in the status bar.
   
Answer:
First, enter a query for the questionable register:


The result is:
2304 = 2^11 + 2^8

2^11 is the sync register.


64 equals 2^6

In the manual, you will find the following information:


If you use our RSSPECAN driver, you can get the register by means of rsspecan_getStatusRegister