Question
The ZVR does not save data in a TOUCHSTONE format.
The ZVR does not save data in a TOUCHSTONE format.
The program TOUCHSTONE requires data from a reflection (S11 or S22) measurement. Or all four S parameters. Then the algorithm will store the data. Please select 'ASCII' under 'items to save'.
The measurement of S12 or S21 alone does not fulfill the TOUCHSTONE requirements and will not be stored.