Using remote commands to enter a label for calibration kit standards

Question

I want to use remote commands to enter calibration kit data. I also want to enter a “label” for the different standards. In the manual I found the command to enter a label for the calibration kit, but not for the standards. What is the remote command?

Answer

In the remote commands, you have to add the label for the standard after the name of the calibration kit.

Here is an example:

CORR:CKIT:N50:MOPEN "ZV-Z123","wolf",0,75600000,0.0189,0,82,0.5,0.09,-0.003

For user-defined connectors, the command is different:

CORR:CKIT:MOPEN '7-16','85038-2','wolf',0,7700000000,0.020,0.00,50,32,0.1,-0.05,0.1,0,0,0,0

In these examples, the “open (m)” standard will be given the label “wolf”.

Thank you.