CMWcards application installed on an external PC can not communicate with the CMW500/290 mobile radio tester

Description

Installation of CMWcards is made on a external PC that should control the CMW mobile radio tester via LAN interface.
But also the PC and the CMW are connected to the same LAN the remote connection do not work and the resource check showed "not connected".

FAQ CMWcards application installed on an external PC - screen 1

Resolution

To use CMWcards with a remote CMW you have to take care that CMWcards and Project Explorer CMW-KT010 (SOAP) is installed on the remote instrument.
The Project Explorer has to be started on the remote CMW to enable the communication via SOAP interface.

FAQ CMWcards application installed on an external PC - screen 2

CMWcards and the Project Explorer tool have to be installed via the Intallation Manager tool.