Question
How can a ZVL with ZVL-K1 option be used instead of an FSL in Remote Control mode so
that a control program such as EMC32 will not recognize this?
How can a ZVL with ZVL-K1 option be used instead of an FSL in Remote Control mode so
that a control program such as EMC32 will not recognize this?
At the start of a program the Remote Control program checks all activated instruments for their status.
This check is started by the *IDN? query.
If instead of an FSL a ZVL with the ZVL-K1 spectrum analyzer function is used in the system, the response to the *IDN? query would cause an error message and the abortion
of the program.
How can the response:
"Rohde&Schwarz,ZVL-6,1303.6509K06/100007,1.10",
be changed to the response of a ZVL with ZVL-K1:
"Rohde&Schwarz,FSL–6,1300.2502K06,100005,1.80" ?
The ID string can be configured in the System Config dialog or via SETUP –
General Setup – GPIB – More – ID String User
Define your own *IDN? response in the editor dialog box.