  • UNCAL Message on the Screen

    UNCAL Message on the ScreenQuestionI get a red UNCAL message on the screen.Full internal calibration was already performed. The sweep time is set to AUTO.What is wrong?AnswerIf you select a wrong setting regarding the 'RF INPUT AC/DC' for the selected frequency range, you will get an 'UNCAL' message.The input of the ESU is divided into a DC-coupled and an AC-coupled frequency range. The ranges are

  • Time Domain Scan without Autoranging

    Time Domain Scan without AutorangingQuestionWhen Time Domain Scan is ON, it is not possible to use autoranging. Is there anything I should take into consideration without autoranging?AnswerIn general, there is no 100% guarantee for an overload-free measurement even if Autoranging is switched on. For example, there is no overload indication for preamplifiers in front of the preselector (e.g.

  • How to set Limit Lines with Frequency Gaps?

    How to set Limit Lines with Frequency Gaps?QuestionHow can I set limit lines that contain frequency gaps?Up to now, you had to enter a pseudo limit of 200 dB, but that doesn't look nice on the screen.AnswerPlease proceed as follows:Select a limit line and split it up into two parts: one before and one after the frequency gap. Create each line with the "new limit line" softkey (after pressing the "

  • Sending a Remote Control Ccommand Internally

    Sending a Remote Control Ccommand InternallyQuestionI need to send a remote control command to the device, but I do not have a PC available to control the device.Can I do this internally with the device itself?AnswerYes, Go to the device's Windows function Open the Windows Explorer Open the C:\Program Files\Rohde&Schwarz\Network Analyzer\Bin folder Execute the iecwin32.exe program Choose the RSIB

  • Remote Desktop: How can I see all Analyzer Buttons on the PC Screen?

    Remote Desktop: How can I see all Analyzer Buttons on the PC Screen?QuestionI have connected my PC to the analyzer via LAN. The firewall settings are made, I have a remote desktop connection, but I cannot see all buttons of the analyzer on my PC.I even shut down the analyzer function and restarted the analyzer function on the analyzer, but this did not help either.AnswerPlease do not restart the

  • How can I Update my Analyzer via LAN?

    How can I Update my Analyzer via LAN?QuestionHow can I update my analyzer with embedded WIN XP via LAN?Answer- Open a remote desktop session to the analyzer on your PC- Close the analyzer function on the PC with ALT + F4- Open Start=> Settings => Control Panel => Security Center => Windows Firewall => Exceptions- Enable 'File and Printer Sharing'- Start the Windows Explorer, go to Tools, select

  • How to check the internal Calibration Source?

    How to check the internal Calibration Source?QuestionDuring calibration I got the error message 'CAL GEN signal low'.How can I check the internal calibration source?AnswerYou can switch the RF path to the internal calibration source in the SETUP - SERVICE menu(menu for the FSL: SETUP - MORE - SERVICE).The calibration source should be a signal of 128 MHz and -30 dBm (for the FSL 65.8 MHz and -20

  • What is the Standard Password?

    What is the Standard Password?QuestionThe login name is 'instrument', but what is the standard password?AnswerDepending on the device and firmware version there are three possible passwords:- 'instrument'- '123456'- '894129'ESCI:Firmware version < 4.42 'instrument'Firmware version >= 4.42 '123456'ESL:'123456'ESPI:Firmware version < 4.42 'instrument'Firmware version >= 4.42 '123456'ESU:Firmware

  • Firmware Rollback

    Firmware RollbackQuestionI want to update my firmware. Can I also roll back the firmware to my old version before firmware update?AnswerWhen the analyzer firmware is updated, the analyzer will boot and operate exactly as before – in some cases even better. But if you want to return to the old version, you can restore the old analyzer firmware by pressing SETUP =>NEXT => FIRMWARE RESTORE.After you

  • Time Domain Scan without Autoranging

    Time Domain Scan without AutorangingQuestionWhen Time Domain Scan is ON, it is not possible to use autoranging. Is there anything I should take into consideration without autoranging?AnswerIn general, there is no 100% guarantee for an overload-free measurement even if Autoranging is switched on. For example, there is no overload indication for preamplifiers in front of the preselector (e.g.

  • Add Limit Line to existing EMC32 Test to create a new list of critical frequencies

    Add Limit Line to existing EMC32 Test to create a new list of critical frequenciesQuestionI have saved an EMI Test and would like to add a new Limit Line to the existing Test, optionally I would like to create a new list of critical frequencies, based on the new limit line. How can I do this with EMC32?AnswerYes, you can do this. You can load the Test and you can add a new limit to the exisitig

  • ESU/FSQ with Windows XP has "Registry File Failure"

    ESU/FSQ with Windows XP has "Registry File Failure"QuestionI got an error message when starting/booting my ESU / FSQ with Windows XP (registry file failure):It is not possible to enter the ESU / FSQ application. (IMGn1039.jpg) shows the error message.stop: c0000218 (Registry File Failure):the registry cannot load the hive(file):\systemroot\system32\config\SYSTEMor its log or alternate.It is

  • Self Alignment via remote control via TCP/IP => 'CAL?

    Self Alignment via remote control via TCP/IP => 'CAL?QuestionI want to make the Self Alignment via remote control with the command 'CAL?'.Because this takes sometimes a very long time how can I detect the end.In the manual is mentioned to use a SRQ, but this does not work when I use a TCP/IP connectionAnswerYes, the SRQ line is a hardware line with the IEEE bus and this does not work with normal

  • How can I Switch from Remote to Local State via Remote Control?

    How can I Switch from Remote to Local State via Remote Control?QuestionHow can I switch from remote to local state via remote control?AnswerUsing a GPIB connection, you can send a Go To Local (GTL) command.If you use VISA, the command is viGpibControlREN and you have to send the VI_GPIB_REN_DEASSERT_GTLmode.You can also use this solution when you perform programming with drivers.If you use the

  • Correct Step Size for a Stepped EMI Scan?

    Correct Step Size for a Stepped EMI Scan?QuestionWhy is it important to use the correct step size for a stepped EMI scan?AnswerIn the case of EMI measurements, the frequency range is analyzed in frequency steps. An important parameter is the IF bandwidth specified by the standards. The step size for a scan depends on the IF range.The step size is important for the following reasons:- If the step

  • Measure Low Frequency Signals with DC Offset

    Measure Low Frequency Signals with DC OffsetQuestionI want to measure noise at a very low frequency (20 Hz to 100 kHz).The signal has a DC offset, but at the input of the analyzer only 0 V DC are allowed with DC coupling.The DC blocks starts at 7 kHz.How can I measure these signals without destroying my device?AnswerAt these frequencies, an external block capacitor can be used. But a rather high

  • Remote command „CALC:UNIT:POW DBUV” is not accepted

    Remote command „CALC:UNIT:POW DBUV” is not acceptedQuestionWe use the receiver mode. After sending the command “CALC:UNIT:POW DBUV” we get the error “-200,"Execution error; Function not available” back. That command is mentioned in the manual and the “unit” dBuV” should be available. What is wrong?AnswerYou are right - in receiver mode it must be possible to use dBuV. In fact, that is the default

  • FSx doesn't recognize active probe RT-ZS30

    FSx doesn't recognize active probe RT-ZS30Question:We have an issue with our FSx and the RT-ZS30 probe: the FSx does not automatically recognize the probeCould you please help us to solve this blocking issue.Answer:You can work with the RT-ZS30 connected to the FSx as follows:+ The RT-ZA9 provides an interface between the probe's BNC socket and an N socket and also supplies the probe with the

Thank you.

Your request was successfully sent!

We will get in touch with you soon.

Your Rohde & Schwarz Team

Download

Please enter your first name.

Please enter a valid email address.

Please enter your name.

Please enter your company name.

Please enter your location.

Marketing permission
Please enter the characters shown in the image
show captcha

The input was not correct.
International Website