R&S®ESCI EMI Test Receiver

For compliance tests meeting all civil standards from 9 kHz to 3/7 GHz

本製品の製造は終了しています
詳細をご希望の方はこちらまでお問い合わせください。

Key Facts

  • Combination of standard-compliant EMI test receiver and high-quality spectrum analyzer
  • Integrated preselection with selectable 20 dB preamplifier

Brief Description

The R&S®ESCI and R&S®ESCI7 EMI test receivers are standard-compliant measuring receivers for EMC certification measurements in line with commercial standards in the frequency range from 9 kHz to 3/7 GHz.

Features

  • Standard-compliant disturbance measurements
  • Spectrum analysis for lab applications
  • Intuitive operation and powerful EMI analysis functions for fast, accurate results

Specifications

Frequency range
R&S®ESCI 9 kHz to 3 GHz
R&S®ESCI7 9 kHz to 7 GHz

Quick Links

Key Facts

  • Combination of standard-compliant EMI test receiver and high-quality spectrum analyzer
  • Integrated preselection with selectable 20 dB preamplifier
  • Effective analysis of the disturbance spectrum through simultaneous graphical presentation of the disturbance level and emission spectrum around the receive frequency (“mixed-mode”)
  • Time-domain analysis for evaluation of timing behavior of disturbances (e.g. click-rate analysis)
  • Simultaneous measurement of multiple traces for parallel evaluation
  • Fast and reliable measurements using automatic and interactive test routines
  • Predefined transducer factors
  • Library of limit lines for commercial standards

Brief Description

The R&S®ESCI and R&S®ESCI7 EMI test receivers are standard-compliant measuring receivers for EMC certification measurements in line with commercial standards in the frequency range from 9 kHz to 3/7 GHz. The receivers conform to the latest version of the CISPR 16-1-1 basic standard. At the same time, they function as full-featured and powerful spectrum analyzers for lab applications.

Features & Benefits

  • Standard-compliant disturbance measurements
    • Characteristics
      • Total measurement uncertainty 1.0 dB (< 3 GHz) and 1.5 dB (3 GHz to 7 GHz)
      • 1 dB compression +5 dBm
      • Displayed average noise level with preamplifier typ. –153 dBm (10 Hz)
      • Twelve preselection filters up to 7 GHz
      • Resolution bandwidths in line with CISPR 16-1-1
      • Detectors: max./min. peak, quasi-peak, average, CISPR-average, RMS, RMS-average
  • Spectrum analysis for lab applications
    • High-quality spectrum analyzer included
    • Extensive measurement functions and evaluation capabilities for general lab applications
  • Intuitive operation and powerful EMI analysis functions for fast, accurate results
    • User-friendly operation as with all EMI test receivers from Rohde & Schwarz
    • SCAN settings in tabular format
    • Simultaneous measurement of multiple traces for parallel evaluation
    • Fast, reliable measurements using automatic and interactive test routines
    • EMI test receiver and spectrum analyzer in a single box – display in MIXED MODE operation
    • Continuous bargraph display and marker functions for precise measurements
    • Time-domain analysis – oscilloscope function in receiver mode
    • Automatic disturbance voltage measurements using remote-controllable line impedance stabilization networks (LISN) from Rohde & Schwarz
    • Predefined transducer factors
    • Library of limit lines for commercial standards
  • Integrates perfectly into the work environment
    • Easy expansion or upgrading by means of options
      • Scalar network analysis
      • RF level measurement
      • Modulation measurement
    • Extensive documentation and networking capabilities like a PC
    • Computer-controlled measurement of disturbances and automated EMI test routines using the R&S®ES-SCAN and R&S®EMC32 EMI measurement software
    • Always up-to-date with free firmware updates
  • Rugged and environmentally friendly, suitable even for field applications
    • Housing option for outdoor use
      • Shock-absorbing corners, carrying handle, carrying bag
    • AC-independent operation using DC power supply and battery pack (option)
    • Flash disk option for expanded environmental specifications
    • Environmentally friendly design
      • Small number of materials
      • Easy identification of materials through labeling
      • Recyclable housing

Specifications

Frequency range
  R&S®ESCI 9 kHz to 3 GHz
  R&S®ESCI7 9 kHz to 7 GHz
Frequency resolution
    0.01 Hz
Frequency accuracy
  Standard 1 × 10-6
  With R&S®FSP-B4 (OCXO) option 1 x 10-7
External reference frequency
    10 MHz
Measurement time
Receiver mode/scan (per frequency step)   selectable from 33 μs to 100 s
Analyzer mode/sweep time   selectable from 2.5 ms to 16000 s,
zero span from 1 μs to 16000 s
Resolution bandwidths
3 dB bandwidths   10 Hz to 3 MHz in steps of 1/3/10
6 dB bandwidths (EMI)   200 Hz, 9 kHz, 120 kHz (–6 dB),
1 MHz (impulse bandwidth)
Video bandwidth, analyzer mode   1 Hz to 10 MHz in steps of 1/3/10
Preselection   can be turned off in analyzer mode
  R&S®ESCI 11 fixed/continuously tunable filters up to 3 GHz
  R&S®ESCI7 12 fixed/continuously tunable filters up to 7 GHz
Level
Max. RF level (CW) input attenuation ≥ 10 dB 137 dBμV (= 1 W)
Max. pulse energy input attenuation ≥ 10 dB  
  R&S®ESCI 10 mWs (20 µs)
  R&S®ESCI7 1 mWs (10 µs)
Max. pulse voltage input attenuation ≥ 10 dB, 10 µs 150 V
Preamplifier can be switched on/off, 20 dB gain  
Third-order intercept (IP3)
   without preselection/preamplifier
  from 200 MHz to 3 GHz > 7 dBm, typ. 10 dBm
  from 3 GHz to 7 GHz > 10 dBm, typ. 15 dBm
R&S®ESCI7
   with preselection/without preamplifier
  from 200 MHz to 3 GHz > 2 dBm, typ. 5 dBm
1 dB compression f > 200 MHz, RF attenuation 0 dB, without preselection/preamplifier > 5 dBm, nominal
Total measurement uncertainty, 95 % confidence level, < 3 GHz
  Without preselection/preamplifier 0.5 dB
  With preselection/preamplifier 1.0 dB
Displayed average noise level (DANL)
receiver mode, nominal, input attenuation 0 dB, termination 50 Ω
  without preamplifier, 1 MHz, BW = 9 kHz < –4 dBμV
  without preamplifier, 30 MHz to 1 GHz,
BW = 120 kHz
< 6 dBμV
  with preamplifier, 1 MHz, BW = 9 kHz < –14 dBμV
  with preamplifier, 30 MHz to 1 GHz,
BW = 120 kHz
< –4 dBμV
Detectors
  receiver mode max./min. peak, quasi-peak, RMS, average, average with meter time, constant (CAV), RMS‑average (CRMS)
Number of measurement points
Analyzer mode   125 to 8001 (default 501)
Receiver mode Frequency domain max 1 million
  Time domain max 1.44 million
ありがとうございます。

リクエストは正常に送信されました。

当社より追ってご連絡いたします。

ローデ・シュワルツ・チーム

デモを取得する

質問を入力してください。
お名前(ファースト・ネーム)を入力してください。
質問を入力してください。
有効な電子メール・アドレスを入力してください。
会社名を入力してください。
市区町村名を入力してください。
郵便番号を入力してください。

国名を入力してください。
マーケティング・パーミッション

次の手段によってローデ・シュワルツから情報を受信することを希望します。


詳細について

私は、このウェブサイトの出版物に記載されているRohde & Schwarz GmbH & Co. KGおよびRohde & Schwarzの法人および子会社が、
ここで選択した手段 (電子メールまたは郵便メール) を通じて、マーケティングおよび広告目的 (特別キャンペーンや値引きに関する情報など) で、私に連絡することに同意します。その内容は、電子計測、セキュリティ通信、モニタリングおよびネットワークテスト、放送およびメディア、そしてサイバーセキュリティ分野の製品やソリューションを含みますが、上記に限定されるものではありません。

お客様の権利

この同意表明は、件名が"Unsubscribe"の電子メールをnews@rohde-schwarz.com宛に送信することにより、いつでも取り消すことができます。これに加えて、今後の電子メール広告の購読を停止するためのリンクが、送信されるすべての電子メールに記載されています。個人データの使用と取り消し手順の詳細は、「プライバシーステートメント」に記載されています。

画像に表示されている文字を入力してください。
show captcha
入力内容が正しくありません。
Do not enter data in this field
ありがとうございます。

リクエストは正常に送信されました。

当社より追ってご連絡いたします。

ローデ・シュワルツ・チーム

見積りを取得する

質問を入力してください。
お名前(ファースト・ネーム)を入力してください。
質問を入力してください。
有効な電子メール・アドレスを入力してください。
会社名を入力してください。
市区町村名を入力してください。
郵便番号を入力してください。
電話番号を入力してください。

国名を入力してください。
マーケティング・パーミッション

次の手段によってローデ・シュワルツから情報を受信することを希望します。


詳細について

私は、このウェブサイトの出版物に記載されているRohde & Schwarz GmbH & Co. KGおよびRohde & Schwarzの法人および子会社が、
ここで選択した手段 (電子メールまたは郵便メール) を通じて、マーケティングおよび広告目的 (特別キャンペーンや値引きに関する情報など) で、私に連絡することに同意します。その内容は、電子計測、セキュリティ通信、モニタリングおよびネットワークテスト、放送およびメディア、そしてサイバーセキュリティ分野の製品やソリューションを含みますが、上記に限定されるものではありません。

お客様の権利

この同意表明は、件名が"Unsubscribe"の電子メールをnews@rohde-schwarz.com宛に送信することにより、いつでも取り消すことができます。これに加えて、今後の電子メール広告の購読を停止するためのリンクが、送信されるすべての電子メールに記載されています。個人データの使用と取り消し手順の詳細は、「プライバシーステートメント」に記載されています。

画像に表示されている文字を入力してください。
show captcha
入力内容が正しくありません。
Do not enter data in this field
ありがとうございます。

リクエストは正常に送信されました。

当社より追ってご連絡いたします。

ローデ・シュワルツ・チーム

お問い合わせ

質問を入力してください。
お名前(ファースト・ネーム)を入力してください。
質問を入力してください。
有効な電子メール・アドレスを入力してください。
会社名を入力してください。
市区町村名を入力してください。
郵便番号を入力してください。

国名を入力してください。
マーケティング・パーミッション

次の手段によってローデ・シュワルツから情報を受信することを希望します。


詳細について

私は、このウェブサイトの出版物に記載されているRohde & Schwarz GmbH & Co. KGおよびRohde & Schwarzの法人および子会社が、
ここで選択した手段 (電子メールまたは郵便メール) を通じて、マーケティングおよび広告目的 (特別キャンペーンや値引きに関する情報など) で、私に連絡することに同意します。その内容は、電子計測、セキュリティ通信、モニタリングおよびネットワークテスト、放送およびメディア、そしてサイバーセキュリティ分野の製品やソリューションを含みますが、上記に限定されるものではありません。

お客様の権利

この同意表明は、件名が"Unsubscribe"の電子メールをnews@rohde-schwarz.com宛に送信することにより、いつでも取り消すことができます。これに加えて、今後の電子メール広告の購読を停止するためのリンクが、送信されるすべての電子メールに記載されています。個人データの使用と取り消し手順の詳細は、「プライバシーステートメント」に記載されています。

画像に表示されている文字を入力してください。
show captcha
入力内容が正しくありません。
Do not enter data in this field
International Website