How to remote-control the UPV via RS-232-C?

Question

I have a UPV and I can see in the Windows control panel under System and Device Manager that the settings for

COM1 are 9600,8,N,1.

I use these settings but no command is accepted.

Answer

The settings for remote control of the UPV via RS-232-C must be as follows:

Baud rate 19200

Data bits 8

Parity None

Stop bits 1

Flow control Hardware