Vertical system |
Analog channels | | 4, 8 |
Digital channels (opt.) | | 16 |
Analog bandwidth (–3 dB) | R&S®MXO54 with (B24x option) | 350 MHz, 500 MHz, 1 GHz, 2 GHz |
| R&S®MXO58 with (B28x option) | 100 MHz, 200, 350 MHz, 500 MHz, 1 GHz, 2 GHz |
Rise time | at 2 GHz bandwidth | 234 ps |
Input impedance | | 50 Ω ± 1.5 %,
1 MΩ ± 1 % || 12 pF (meas.) |
Input sensitivity | at 50 Ω | 0.5 mV/div to 3 V/div, support entire analog bandwidth |
| at 1 MΩ | 0.5 mV/div to 10 V/div, support entire analog bandwidth |
Vertical resolution | | up to 18 bit with HD mode (high definition) |
DC gain accuracy | input sensitivity > 5 mV/div | ±1 % full scale |
Offset range at 50 Ω | < 33 mV/div | ±(2 V – input sensitivity × position) |
| 33 mV/div to < 120 mV/div | ±(7 V – input sensitivity × position) |
Offset range at 1 MΩ | 0.5 mV/div to < 80 mV/div | ±(5 V – input sensitivity × position) |
| 80 mV/div to < 800 mV/div | ±50 V |
Channel-to-channel isolation | | > 60 dB (1:1000) |
Acquisition system |
Sampling rate | analog channels (real time) | max. 5 Gsample/s on 2 channels,
max. 2.5 Gsample/s on 4 channels |
| analog channels (interpolated) | max. 5 Tsample/s |
| digital channels | max. 5 Gsample/s on each channel |
Acquisition memory | Standard: R&S®MXO58, 8 channel instrument | max 500 Mpoints with 8 active channels (single capture)
max. 500 Mpoints with 4 active channels (run continuous) |
| Standard:R&S®MXO54, 4-channel instrument: | max. 500 Mpoints (single and run continuous) |
| Optional R&S®MXO5-B110 memory option 1 Gpoints | max. 1 Gpoints with 4 active channels (single capture)
max. 1 Gpoints with 2 active channels (run continuous) |
History segments | ≤ 1 Mpoints record length | 1 million segments |
Max. acquisition rate | | > 4 500 000 waveforms/s |
Acquisition modes | | sample, peak detect, average, envelope |
Interpolation modes | | linear, sin(x)/x, sample & hold |
Horizontal system |
Timebase range | | Selectable between 200 ps/div and 10 000 s/div |
Timebase accuracy | after delivery/calibration, at +23 °C | ±0.2 ppm |
| During calibration interval | ±1 ppm |
Channel-to-channel skew | between analog channels | < 100 ps (meas.) |
| between digital channels | < 500 ps (meas.) |
Trigger system |
Standard trigger types | | Edge, glitch, width, runt, window, timeout, interval, slew rate, setup&hold, pattern, state |
Advanced trigger modes | | A/B/R, serial bus, trigger in and out |
Trigger sensitivity | | 0.0001 div, from DC to instrument bandwidth for all
vertical scales |
Analysis and measurement |
Automatic measurements | | 39 |
Cursor measurements | | 4 |
Waveform mathematics | | Up to 8 equations math on math (add, subtract, multiply, divide, absolute value,
square, square root, integrate, differentiate,
log10, loge, log2, reciprocal, invert, low pass,
high pass, rescale (a*x+b)) |
Logic analyzer (MSO) functionality |
Digital input channels | | 16 |
MSO Bandwidth | | 400 MHz |
MSO sampling rate | | 5 Gsamples/s per channel |
General data |
Dimensions | W × H × D | 445 mm × 314 mm × 153 mm
(17.51 in × 12.36 in × 6.02 in) |
Weight | without options, nominal | 9.0 kg (19.85 lb) |
Rackmount height | | 8 HU |
I/O Interfaces | Front | 3 × USB 3.1 Gen 1 ports, type A plug |
| Rear | 1 × USB 3.1 Gen 1 port, type B plug |
| Right | 2 × USB 3.1 Gen1 ports, type A plug
RJ-45 connector, supports 10/100/1000BASE-T
HDMI™ 2.0 and DisplayPort++ 1.3, output of oscilloscope display |
Display | type | 15.6" LC TFT color display with capacitive
touchscreen |