Question:
I am using the trigger type "EDGE A/B".
How can I set the levels for the A and for the B trigger with remote commands?
The command " TRIG:A:LEV ..xx" does not work for that trigger type.
Question:
I am using the trigger type "EDGE A/B".
How can I set the levels for the A and for the B trigger with remote commands?
The command " TRIG:A:LEV ..xx" does not work for that trigger type.
Answer:
Here are the commands to set the levels:
TRIG:A:LEV:RIS:LOW 4.2500E-01 --> level for A Trigger (set to 425 mV )
TRIG:A:LEV2:RIS:LOW 2.00000E-01-->level for B Trigger (set to 200 mV )