FAQs from Rohde & Schwarz

Correction Column in EMC32-EB Final Results

Question

In using EMC32-EB with EMC32-K10 EMI Auto Test.This option expands our EMC32-EB to provide test sequences for fully automatic testing. At the end of the Test, we get a Final Result table. Is it possible to see individual losses of cable, antenna, pre-amp, etc in the final table instead of a single total correction value?

Answer

Yes, this is possible. You can influence the tables in multiple ways. Here are some ways:

At first, in EMI Auto Test there is a way to customize the tables from the General Settings. The requested functionality is also available from here.

Correction Column in EMC32-EB Final Results

Please use this feature "Detailed Measurement Correction Information" if you want to get detailed information on the measurement correction in your tables. In this case new columns for contribution of signal path, antenna / transducer cable, antenna / transducer factor and the raw measurement value will be added to the Critical and Final Result table. The table looks like the following example:

Correction Column in EMC32-EB Final Results

Beside this, you have more options from the General Settings menu. You may also Generate Final Result Table for each Detector or Add Time/Date of Measurement to Comment Column.

In addition, the Interactive data reduction is able to interrupt the test after the data reduction has been completed in order to edit the list points to be edited or expanded. There is one more interactive data reduction to revise the results from the latest test phase (typically from the maximization) before the final measurements.

Correction Column in EMC32-EB Final Results

This allows you to delete frequencies from the table with a right mouse click

Correction Column in EMC32-EB Final Results

Also the header of the table allows changes. With the corresponding icons of the table header it is possible to add new frequencies to the table, import crtiticsl frequencies from a table or to import critical frequencies from another test.

Correction Column in EMC32-EB Final Results

As you can see, you have many possibilities to adjust the table to your needs.