Question

How can we find out if calibration is switched on or off?

(as it is indicated in the display)

Answer

The [SENSe[1...4]:]CORRection:STATe ON | OFF command switches

calibration on or off.

The [SENSe[1...4]:]CORRection:STATe? query will return 1 if calibration

is switched on, and it will return 0 if calibration is switched off.