DC Source/Measurement |
Source/measurement of DC voltage/current |
Pulse Source/Measurement |
Source/measurement of pulse voltage/current (However, measurement auto range in pulse source is impossible.) |
DC Sweep Source/Measurement |
Source/measurement with linear, two-slope linear, random and fixed level. |
Pulse Sweep Source/Measurement |
Source/measurement with linear, two-slope linear, random and fixed level. (However, measurement auto range in pulse source is impossible.) |
Low-Resistance Measurement |
By pulse current source / voltage measurement |
Integration Time |
100µs, 500µs, 1ms, 5ms, 10ms, 1PLC, 100ms, 200ms, or S/H S/H: Measuring in Sample & Hold Mode (Integration time: 100µs), (Enabled only in pulse source and pulse sweep source modes.) PLC: Power Line Cycle: 50Hz:20ms 60Hz:16.66ms |
Sweep Mode |
Reverse ON/OFF (forward to backward / one way) |
Sweep Repeat Time |
1 to 1000 times, or infinite |
Number of Sweep Maximum Step |
8000 steps |
Random Sweep Maximum Memory |
8000 data |
Measurement Data Memory |
8000 data |
Measurement Auto Range |
Available only in VSIM or ISVM |
Measurement Function Link Mode |
Links the source function to the measurement function (VSIM or ISVM) ON/OFF available |
Limiter |
Possible to set separately at HI or LO sides (However, in case of current limiter, the same polarity can not be set.) |
Calculation Functions |
NULL calculation Comparator calculation (HI/GO/LO) Scaling calculation MAX/MIN/AVE/TOTAL calculation |
Trigger Style |
Auto-Trigger, external Trigger |
Output Terminal |
Front; Safety sockets, HI OUTPUT, HI SENSE, LO OUTPUT, LO SENSE |
Maximum Input Voltage Between Terminals |
15V peak MAX (between HI and LO) 2V peak MAX /(between OUTPUT and SENSE) 250V MAX (between LO and chassis) |
Maximum Remote Sensing Voltage |
± 1V MAX between HI OUTPUT and HI SENSE between LO OUTPUT and LO SENSE (The voltage between HI SENSE and LO SENSE must be within the maximum output voltage range) |
Voltage Measurement Input Resistance |
1 GΩ or higher |
Voltage Measurement Input Leakage Current |
± 1 nA or lower |
GPIB Interface |
Compliant with IEEE-488.2-1987 Interface functions: SH1, AH1,T5, L4, SR1, RL1, PP0, DC1, DT1, C0, E2 Connector: Amphenol 24 pin |
USB Interface |
USB 2.0 Full-Speed; Connector Typ B |
External Control Signal |
TRIGGER IN, INTERLOCK, OPERATE IN/OUT, COMPLETE OUT, SYNC OUT |