Question
My HMC8012 shows slightly to high resistance / current when preparing for a measurement task. I tried to perform a zero calibration using the softkey "Null", but this did not work at all. Is there another way to perform this step?
My HMC8012 shows slightly to high resistance / current when preparing for a measurement task. I tried to perform a zero calibration using the softkey "Null", but this did not work at all. Is there another way to perform this step?
In terms of zero calibration the hard key "NULL" should be used instead of the soft key.
The softkey allows to change the offset manually, but does not automatically use the current correction value as adjustment variable. Please prepare your measurement with the following steps:
1. Resistance
2. Current
Prerequisites: HMC8012 with FW 1.400 or later, adequate warm up time (90 minutes) recommended.