How to Connect the Control Cable from the FSP-B10 (ext. Generator Control) to the SMU200A?

Question

I have a spectrum analyzer with the external generator control option FSP-B10 and I want to connect the control cable to the SMU200A. Which BNC connector on the SMU200A do I have to use for 'Marker', 'Blank' and 'Trigger'?

Answer

Connect the 'Blank' connector of the cable to the 'User1' connector of the SMU200A.

Connect the 'Trigger' connector of the cable to the 'INSTR TRIG' connector of the SMU200A.

The 'Marker' connector of the cable is not connected to the SMU200A.

Use the same connectors on the SMATE and SMJ100A.