ZVA - One Path two Port Calibration with Port 3 and 4

Question:

How can I configure a "One Path two Port Calibration" for Port 3 and 4. It can not be found in the calibraiton menu.

Answer:

For such a calibration you have to enter the "Calibration" menu:

At "Start Cal" select "Other"

Select the connector type, gender and a suitable calibration kit, then click on "Next"

A new window opens:

Rohde-Schwarz-FAQ-ZVA-One-Path-two-Port-Calibration-with-Port-3-and-4_screen02_5.jpg
ZVA - One Path two Port Calibration with Port 3 and 4

Click on "Add Calibration"

At the drop down menu select "One Path Two Port"

Set the check mark at the ports to be calibrated

Finally set the "Node Port"

Now you can start the calibration