Interested?

R&S®Essentials | Power supplies fundamentals

Understanding output response

Author: Anja Fenske, Product Manager R&S Essentials Power supplies

It is important to choose a power supply that can follow quickly changing load conditions. If, for example, a device switches from very low to high current consumption, the power supply will take some time before its output reaches its final setting. On the other hand, power supplies can react with overshoots that endanger sensitive DUTs.

Depending on the type of power supply, it will respond to abrupt load changes with slower or faster recovery times.

Load recovery time

Load recovery time, also referred to as transient response, describes the time required by a DC power supply to reach a set voltage level after a load change. The power supply’s output voltage drops when the current suddenly rises. Likewise, a decrease in current causes a momentary increase in the power supply’s voltage. Load recovery time is the duration after which a power supply has recovered from such transient caused by a significant load change.

Consumer electronics such as mobile phones and IoT devices require very little power in sleep mode. However, the current increases abruptly when the device switches to transmit mode. A power supply used to power such DUTs must be capable of handling load changes from a few nA to the ampere range with minimum voltage drops and overshoots.

Specialty power supplies have optimized control circuits that react very fast to load changes and minimizeovershoots.

The R&S®NGL/NGM series power supplies and the R&S®NGU source measure units allow users to choose how the instrument should respond to load changes.

The “Fast” default setting is optimized for speed, achieving recovery times of < 30 μs. Deactivating “Fast” will slightly increase recovery time, focusing on preventing overshoots under special load conditions.

Voltage priority and current priority mode

Setting and regulating the output voltage (constant voltage mode) is the standard mode for power supplies. Most power supplies can also be used in constant current mode, where current limiting ensures that only the configured current can flow. However, these devices are not optimized for fast current limiting. There is a risk of damage to sensitive DUTs due to excessive currents from overshoots in current regulation. To avoid this risk, the R&S®NGU source measure units have separate operating modes for voltage and current regulation.

Curious to learn more about test fundamentals?

Sign up for our newsletter

Further educational content