EMC32, add CDNE Coupling-/ Decoupling Network for EMI measurements

Question

I want to add a CDNE Coupling-/ Decoupling Network into EMC32 for EMI measurements. I would like to know how to add CDNE to Device List to create a Hardware Setup.

Answer

At first you have to add a device of the type Transducer to EMC32 Device List.

EMC32, add CDNE Coupling-/ Decoupling Network for EMI measurements

It makes sense to rename the device properly:

EMC32, add CDNE Coupling-/ Decoupling Network for EMI measurements

The new device you have to be configured as an EMI clamp (without positioning device):

EMC32, add CDNE Coupling-/ Decoupling Network for EMI measurements

Don't forget to add the correction table for this device:

EMC32, add CDNE Coupling-/ Decoupling Network for EMI measurements

Note: If the Correction Table is not yet available, it must first be made available to EMC32 from the calibration data of the manufacturer:

EMC32, add CDNE Coupling-/ Decoupling Network for EMI measurements

Now, CDNE can be inserted into an EMI Conducted Hardware Setup.

EMC32, add CDNE Coupling-/ Decoupling Network for EMI measurements

Based on this you create a Test Template and start a new Test.