GPIB sequence for performing the "normalization" calibration method
Question
I am looking for a GPIB sequence for performing the "normalization" calibration method with the R&S ZVL at port 1. For
calibration I want to use the SHORT from the R&S ZV-Z121 calibration kit. The calibration kit is already available on the R&S ZVL:
I need to include the calibration in a measurement automation program. Do you have an example?
Answer
Yes, it is possible to perform this measurement with GPIB commands. We recommend the following sequence:
ZVL: *RST
ZVL: SYSTEM:DISPAY:UPDATE ON
ZVL: INST:NSEL 2
ZVL: CALC:PAR:MEAS 'Trc1','S11'
ZVL: SENSE:CORRECTION:CKIT:N50:SELECT 'ZV-Z121'
ZVL: SENSE:CORRECTION:COLLECT:CONNECTION1 N50MALE
ZVL: SENSE:CORRECTION:COLLECT:METHOD:DEFINE 'TEST SFK NORM PORT1', RSHORT, 1
ZVL: SENSE:CORRECTION:COLLECT:ACQUIRE:SELECTED SHORT, 1
ZVL: SENSE:CORRECTION:COLLECT:SAVE:SELECTED