Results 3
  • Remote command „CALC:UNIT:POW DBUV” is not accepted

    We use the receiver mode. After sending the command “CALC:UNIT:POW DBUV” we get the error “-200,"Execution error; Function not available” back. That command is mentioned in the manual and the “unit” dBuV” should be available. What is wrong?You are right - in receiver mode it must be possible to use dBuV. In fact, that is the default setting after a preset.Please check if there is any other transducer

  • Add Limit Line to existing EMC32 Test to create a new list of critical frequencies

    I have saved an EMI Test and would like to add a new Limit Line to the existing Test, optionally I would like to create a new list of critical frequencies, based on the new limit line. How can I do this with EMC32?Yes, you can do this. You can load the Test and you can add a new limit to the exisitig graphic. It is also possible to run a new EMI Evaluation to get new critical frequencies. Attached

  • Self Alignment via remote control via TCP/IP => 'CAL?

    I want to make the Self Alignment via remote control with the command 'CAL?'.Because this takes sometimes a very long time how can I detect the end.In the manual is mentioned to use a SRQ, but this does not work when I use a TCP/IP connectionYes, the SRQ line is a hardware line with the IEEE bus and this does not work with normal TCP/IP. So we cannot use it.Because the calibration command is a query

Thank you.

Your request was successfully sent!

We will get in touch with you soon.

Your Rohde & Schwarz Team

Download

Please enter your first name.

Please enter a valid email address.

Please enter your name.

Please enter your company name.

Please enter your location.

Marketing permission
Please enter the characters shown in the image
show captcha

The input was not correct.
International Website