Frequency
|
Frequency range |
R&S®FPL1003 |
5 kHz to 3 GHz |
|
R&S®FPL1007 |
5 kHz to 7.5 GHz |
Frequency resolution |
|
0.01 Hz |
Bandwidth
|
Resolution bandwidth (–3 dB) |
sweep filters |
100 kHz to 10 MHz in 1/2/3/5 sequence |
|
FFT filters |
1 Hz to 50 kHz in 1/2/3/5 sequence |
Signal analysis bandwidth (equalized) |
standard |
nom. 10 MHz |
|
with R&S®FPL1-B40 option |
nom. 40 MHz |
Displayed average noise level (DANL)
|
RF preamplifier off |
5 kHz ≤ f < 100 kHz |
typ. –143 dBm |
|
100 kHz ≤ f < 5 MHz |
< –140 dBm, typ. –143 dBm |
|
5 MHz ≤ f < 3 GHz |
< –149 dBm, typ. –152 dBm. |
|
3 GHz ≤ f < 5 GHz |
< –143 dBm, typ. –146 dBm |
|
5 GHz ≤ f ≤ 7.5 GHz |
< –140 dBm, typ. –143 dBm |
RF preamplifier on (gain nom. 20 dB) |
3 MHz ≤ f < 10 MHz |
< –155 dBm, typ. –158 dBm |
|
10 MHz ≤ f < 2 GHz |
< –163 dBm, typ. –166 dBm |
|
2 GHz ≤ f < 3 GHz |
< –162 dBm, typ. –165 dBm |
|
3 GHz ≤ f < 5 GHz |
< –158 dBm, typ. –161 dBm |
|
5 GHz ≤ f < 7 GHz |
< –156 dBm, typ. –159 dBm |
|
7 GHz ≤ f < 7.5 GHz |
< –155 dBm, typ. –158 dBm |
Attenuation and Gain
|
Setting range of attenuator |
standard |
0 dB to 45 dB, in 5 dB steps |
|
with R&S®FPL1-B25 option |
0 dB to 45 dB, in 1 dB steps |
PRF preamplifier gain |
with R&S®FPL1-B22 option |
nom. 20 dB |
Intermodulation
|
1 dB compression of input mixer (two-tone) |
RF attenuation 0 dB, RF preamplifier = off |
nom. +7 dBm |
Third-order intercept point (TOI) |
RF attenuation 0 dB, level 2 × –20 dBm, ∆f > 5 × RBW or 10 kHz, whichever is larger, RF preamplifier = off |
|
|
10 MHz ≤ fin < 300 MHz |
> 13 dBm, typ. 16 dBm |
|
300 MHz ≤ fin < 3 GHz |
> 17 dBm, typ. 20 dBm |
|
3 GHz ≤ fin ≤ 7.5 GHz |
> 15 dBm, typ. 18 dBm |
|
with R&S®FPL1-B22 option, RF attenuation 0 dB, level 2 × –40 dBm, ∆f > 5 × RBW or 10 kHz, whichever is larger, RF preamplifier = on |
|
|
5 MHz ≤ fin < 7,5 GHz |
nom. 0 dBm |
Phase noise
|
SSB phase noise |
frequency = 1000 MHz, carrier offset |
|
|
100 Hz |
nom. –88 dBc (1 Hz) |
|
1 kHz |
< –99 dBc (1 Hz) |
|
10 kHz |
< –105 dBc (1 Hz), typ. –108 dBc (1 Hz) |
|
100 kHz |
< –110 dBc (1 Hz), typ. –115 dBc (1 Hz) |
|
1 MHz |
< –130 dBc (1 Hz), typ. –135 dBc (1 Hz) |
|
10 MHz |
nom. –152 dBc (1 Hz) |
Overall Amplitude Accuracy
|
Total measurement uncertainty |
signal level 0 dB to –50 dB below reference level, S/N > 20 dB, sweep time auto, sweep type = FFT, RF attenuation 10 dB, 20 dB, 30 dB, 40 dB, RF preamplifier = off, span/RBW < 100, confidence level 95 %, +20 °C to +30 °C |
0.5 dB |
|
1 MHz ≤ f < 3 GHz |
0.5 dB |
|
3 GHz ≤ f < 7.5 GHz |
0.8 dB |