Description
Using my Turntable the Final Adjustment works on full range (from -180 to +180°) , also if I have defined a "predefined range" of 60°-
Using my Turntable the Final Adjustment works on full range (from -180 to +180°) , also if I have defined a "predefined range" of 60°-
This may happen and this function is intended. For better understanding, here some background information:
1. The turntable range from -180 to +180 deg does not support any additional range, e.g. -210 to + 210.
2. For example, assume our critical frequency from preview or zoom measurement is found @ -160 deg.
3. If you have set a search range of 60 deg, thus software needs to search from -160 deg - 30 deg = -190 deg to -160 + 30 deg = -130 deg.
4. Since the min angle is set to -180 the adjustment algorithm needs to measure in the ranges -180 deg to -130 deg and 170 deg to 180 deg to cover this search range.
5. Since it would take more time to stop the measurement at -130 deg, move to 170 deg and restart the measurement for only 10 deg it is hard coded that in this case we do a full range measurement. --> this has no disadvantages for the result.
There is a solution for you to avoid this.
a) Activate the "range clippling" here in the accessories settings. --> search range is reduced
b) extend the turntable range for additional 30 deg on both sides if the Hardware is capable to do this.