HM6050-2 technical differences between EU/GB/EU Versions

Question

Is it possible to run a EU version 6050 (230V) in a 115V network?

Answer

You can use the EU version in a 115V network and will have valid measurement values. The only differences between the units are

  • the power cord (Input)
  • the power plug (Output)
  • the setting of the (solder-) bridges on the power supply board

When e.g. the EU Version is used in the american network, you can do all the measurements as intended. The only restriction is, that you won't be able to switch from L1 to N.