Question
How can I get the Peak Hold value of a single measurement in Receiver mode? Is there a remote control Sequence for this?
How can I get the Peak Hold value of a single measurement in Receiver mode? Is there a remote control Sequence for this?
Below you will find a command sequence as an example. As you can see I configure bargraph in Receiver Mode for 500 MHz with detector Quasipeak and I do 4 measurements.
After this I do a query on the last single measurement result and on the Bargraph MaxHold result.
*IDN?
*RST
*CLS
FORMAT ASCII
INSTRUMENT RECEIVER
FREQUENCY:CW 500MHZ
DETECTOR:RECEIVER QPEAK
INIT:CONT OFF
DISP:BARG:PHOL ON
INIT
INIT
INIT
INIT
TRACE? SING
TRACE? PHOLD
As you can see, with INIT:CONT OFF you change to the single measurement and with DISP:BARG:PHOL ON the bargraph display is activated..Each INIT starts a single measurement. In my case 4 single measurements are done. TRACE? SING reads out the last (current) single measurement result, the command TRACE? PHOLD reads the bargraph MaxHold value of all four measurements.. Each PHOLD result is made up out of two values, absolute level and frequency.