For correct detection of the signal it is necessary to select a faster timebase.
The attached screenshots shows an example. The signal is a 13 MHz CW with pulse modulation.
SCR10.PNG: Acquire Mode "Sample" --> everything is correct
SCR9.PNG: same time base as before with "High Resolution" --> the sample rate is too low to detect the 13 MHz signal correct
SCR11.PNG: "High Resolution" with faster time base. The amplitude is almost correct.
Conclusion:
For the Acquire Mode "High Resolution" a faster time base is necessary!