FAQs from Rohde & Schwarz

ZVH: Store the measurement as S2P Touchstone file

Question

How can I store the measurement result of the ZVH as a S2P file in touchstone format?

Answer

You can store S2P files from the ZVH via the help of the software ZVHView.

Here the link to the ZVHView:
https://www.rohde-schwarz.com/software/zvh8/

Do the following:

  • make a Full-2-Port calibration on the ZVH
  • do the measurement
  • switch to the ZVHView and connect to the ZVH
  • Display Sweep under Instrument.
  • save the measurement result under File - Save As in S2P file format: