FAQs from Rohde & Schwarz

The measurement range does not fit using a sensor

FAQ-Compensation-type-symbols-in-main-display-of-the--LCX_LCR-Meter_Screenshot8.png

Question:

I set up my HMC8015 for current measurements using a shunt.

I do not understand why the calculated current range does not fit to my settings.

I performed all the settings in the [EXT] - (Sensor) menu and would expect it to calculate a maximum current of 4 V / 100 mOhm = 40 A.

However, it shows 13.3 A. What is the reason for this behavior?

FAQ-Compensation-type-symbols-in-main-display-of-the--LCX_LCR-Meter_Screenshot9.png

Answer:

The displayed current is RMS based, while your calculation example only covers peak values.

The HMC8015 allows to set two different values for the crest factor:

FAQ-Compensation-type-symbols-in-main-display-of-the--LCX_LCR-Meter_Screenshot10.png

In terms of current the crest factor k(c) is calculated using the peak value I(p) and the RMS value I(rms):

k(c) = I(p) / I(rms)

In other words: When the measurement range is based on a crest factor k(c) = 3 and the peak value I(p) is 40 A, the

RMS value I(rms) is

I(rms) = I(p) / k(c) = 40 A / 3 = 13.33 A

FAQ-Compensation-type-symbols-in-main-display-of-the--LCX_LCR-Meter_Screenshot11.png