Input channels |
R&S®RTO1002/R&S®RTO1012/ R&S®RTO1022 |
2 |
|
R&S®RTO1004/R&S®RTO1014/ R&S®RTO1024/R&S®RTO1044 |
4 |
Analog bandwidth (–3 dB) at 50 Ω |
R&S®RTO1002 and R&S®RTO1004 |
600 MHz |
|
R&S®RTO1012 and R&S®RTO1014 |
1 GHz |
|
R&S®RTO1022 and R&S®RTO1024 |
2 GHz |
|
R&S®RTO1044 |
4 GHz |
Rise time (calculated) |
R&S®RTO1002 and R&S®RTO1004 |
583 ps |
|
R&S®RTO1012 and R&S®RTO1014 |
350 ps |
|
R&S®RTO1022 and R&S®RTO1024 |
175 ps |
|
R&S®RTO1044 |
100 ps |
Input impedance |
|
50 Ω ±1.5 % 1 MΩ ±1 % with15 pF (meas) |
Input sensitivity |
max. bandwidth in all ranges |
50 Ω: 1 mV/div to 1 V/Div 1MΩ: 1 mV/div to 10 V/Div |
ENOB of the A/D converter |
full-scale sine, frequency < –3 dB bandwidth |
> 7 bit (meas) |
Acquisition system
|
Realtime sampling rate |
R&S®RTO1002/R&S®RTO1004/ R&S®RTO1012/R&S®RTO1014/ R&S®RTO1022/R&S®RTO1024 |
10 Gsample/s on each channel |
|
R&S®RTO1044 |
max. 10 Gsample/s on 4 channels, max. 20 Gsample/s on 2 channels |
Memory depth |
standard configuration, per channel/1 channel active |
R&S®RTO2000: 2-channel-model: 20 /40 Msample R&S®RTO2000: 4-channel-model 20 /80 Msample |
|
max. upgrade (R&S®RTO-B102 option), per channel/1 channel active |
R&S®RTO2000: 2-channel-model: 100 /200 Msample R&S®RTO2000: 4-channel-model: 100 /400 Msample |
Max. acquisition rate |
continuous aquisition and display, 10 Gsample/s, 1 ksample |
1 000 000 waveforms/s |
|
ultra-segmented-mode |
< 300 ns blind time |
Decimation modes |
any combination of decimation mode and waveform arithmetics on up to 3 waveforms per channel |
Sample, Peak Detect, High Resolution, Root Mean Square |
Waveform arithmetics |
|
off, envelope, average |
Interpolation modes |
|
Linear, Sin(x)/x, Sample&Hold |
Horizontal system
|
Time base range |
|
25 ps/div to 10000 s/div |
Time base accuracy |
|
±5 ppm |
|
after delivery/calibration R&S®RTO-B4 option |
±0.02 ppm |
Channel deskew |
|
±100 ns (realtime deskew, channel-to-channel triggers - e.g. State - detect deskew) |
Trigger system
|
Trigger types |
|
edge, glitch, width, runt, window, timeout, interval, slew rate, data2clock, pattern, state, serial pattern, TV/video, serial bus trigger (optional), SENT |
Sensitivity |
definition of trigger hysteresis |
can be set autimatically or manually from 0 div to 5 div |
Min. detectable glitch |
R&S®RTO1002/R&S®RTO1004/ R&S®RTO1012/R&S®RTO1014/ R&S®RTO1022/R&S®RTO1024 |
100 ps |
|
R&S®RTO1044 |
50 ps |
Waveform maths
|
Algebraic categories |
|
mathematics, logical operations, comparison, frequency domain, digital filter |
Hardware-accelerated mathematics |
+, -, *, 1/x, |x|, derivative, log10, ln, log2, scaling, FIR, FFT magnitude |
Analysis and measurement functions
|
Hardware-accelerated analysis |
|
spectrum, histogram, mask test, cursor |
Hardware-accelerated measurements |
|
amplitude measurements, time measurements |
General data
|
Dimensions |
W x H x D |
427 mm x 249 mm x 204 mm (16.81 in x 9.8 in x 8.03 in) |
Weight |
R&S®RTO1024 |
9.6 kg (21.16 lb) |
Display |
|
10.4" LC TFT color touch screen, 1024 x 768 pixels (XGA) |
Connectivity |
|
1 Gbit/s LAN, 4 x USB 2.0, GPIB (optional), DVI for external monitor, external trigger, trigger output |