R&S®SMW-K501 Extended Sequencing

  • Übersicht

Kerndaten

  • Sequencing of arbitrary waveform segments
  • Realtime change of amplitude, offset frequencies, relative phase, pulse width and off times
  • Support of loops, nested loops and repetitions

Kurzbeschreibung

The R&S®SMW-K501 extended sequencing option can be used manually via sequencing lists and waveform segments or via the R&S®Pulse Sequencer Software and its options R&S®SMW-K300 and R&S®SMW-K301.

Technische Daten

Specifications
Modes User or R&S®Pulse Sequencer software
User mode sequencing via user-defined XML lists
R&S®Pulse Sequencer Software mode sequencing via R&S®Pulse Sequencer Software

Kerndaten

  • Realtime signal generation of CW, unmodulated rectangular CW pulses and pulses with linear frequency modulation or Barker codes with the R&S®Pulse Sequencer Software together with the options R&S®SMW-K300 and R&S®SMW-K301
  • Sequencing of arbitrary waveform segments
  • Realtime change of amplitude, offset frequencies, relative phase, pulse width and off times
  • Signal generation based on time of arrival information (ToA) embedded into the sequencing list
  • Ultralong playtimes
  • Support of loops, nested loops and repetitions
  • The R&S®SMW-K501 extended sequencing option can be used manually via sequencing lists and waveform segments or via the R&S®Pulse Sequencer software and options R&S®SMW-K300 and R&S®SMW-K301
  • Intuitive and easy-to-use graphical user interface in manual mode

Kurzbeschreibung

The R&S®SMW-K501 extended sequencing option can be used manually via sequencing lists and waveform segments or via the R&S®Pulse Sequencer Software and its options R&S®SMW-K300 and R&S®SMW-K301.

If used with the R&S®Pulse Sequencer Software and its options, the software calculates all required signals and required sequencing lists and transfers them to the signal generator. Continuous wave signals, unmodulated rectangular CW pulses and pulses with linear frequency modulation or Barker codes are calculated in realtime inside the R&S®SMW200A. Also pre-calculated waveform segments can be used with these sequencing lists. Changes of amplitude, offset frequency, pulse width, offset phase and off time are always applied in realtime as defined by the sequencing list. The start time of the signal generation process is defined by a relative time stamp with time of arrival (ToA) information. In user mode, the R&S®SMW-K501 option allows sequencing of waveform segments by using user defined XML sequencing lists. It supports loops, nested loops and repetitions, enabling ultralong playtimes. In both cases, memory requirements are reduced to a minimum and playtime is increased enormously.

Technische Daten

Specifications
Modes  
User modesequencing via user-defined XML lists  
R&S®Pulse Sequencer Software modesequencing via R&S®Pulse Sequencer Software 
Parameter types for R&S®Pulse Sequencer Mode
Data format
Sequencing file mandatory, memory shared with I/Q data and segment addresses 
File size
Sequencing file rectangular pulses, unmodulated, variable amplitude, 5 ms PRI187 kbyte/min
 rectangular pulse, linear FM, variable FM, variable amplitude, phase and frequency, 5 ms PRI305 kbyte/min.
Setting granularity
Time with R&S®SMW-B10 option5 ns
Amplitude 16 bit (voltage-based)
Phase <0.01 degrees
Frequencywith R&S®SMW-B10 option 0.05 Hz
I/Q segments
Maximum individual segments 16 777 216 
Length granularitywith R&S®SMW-B10 option32 samples
Timing
Maximum play timewith R&S®SMW-B10 option24 h
Minimum pulse widthwith R&S®SMW-B10 option, real time1 sample, 5 ns
 with R&S®SMW-B10 option, I/Q segment1 samples, 5 ns
Minimum PRI 1 µs
Frequency switchingwith R&S®SMW-B10 option1 sample, 5 ns
Real time MOP types
Linear FM up, down, triangular
Phase Barker
Modes
Linear FM up, down, triangular
Phase Barker