Is my ESR or ESW ready for CISPR 16-1-1:2019 (Ed.5)?

Description

I already have ESR/ESW Receivers and I would like to know, if this existing device can be calibrated as required by CISPR 16-1-1:2019 (Ed.5)?

Resolution

It is possible get the ESW and ESR ready for CISPR 16-1-1:2019 (Ed.5).:

  • At first, make sure that the instrument is "Edition 4 Ready". (Please open from menu: Setup => System Config => Hardware Info)
  • Update of instrument to latest firmware is recommended
  • Register instrument for calibration at R&S Service Department and for On-site Calibration
  • After calibration, “CISPR 16-1-1 Edition 5 Ready" is displayed on instrument. (Please open from menu: Setup => System Config => Hardware Info)